Physical Dosimetry in an Anthropomorphic Phantom Reveals Superior Organ Sparing with Multicatheter Interstitial Brachytherapy Over MammoSite for APBI
World Journal of Nuclear Medicine
A rigorous phantom dosimetry study with direct physical measurements showing MIB superiority in OAR sparing over MBS, but limited to in vitro conditions and requiring confirmation in patient-specific studies.
Reported
Heart dose MBS30.9% prescribed dose
Heart dose MIB19.1% prescribed dose
Ipsilateral lung dose MBS23.5% prescribed dose